![]() ![]() Spectrophotometric tests contribute meaningfully toward the identification of many compendial chemical substances. The test procedures that follow are applicable to substances that absorb IR and/or UV radiation (see Spectrophotometry and Light-Scattering
![]() ![]() The IR absorption spectrum of a substance, compared with that obtained concomitantly for the corresponding USP Reference Standard, provides perhaps the most conclusive evidence of the identity of the substance that can be realized from any single test. The UV absorption spectrum, on the other hand, does not exhibit a high degree of specificity. Conformance with both IR absorption and UV absorption test specifications, as called for in a large proportion of compendial monographs, leaves little doubt, if any, regarding the identity of the specimen under examination.
INFRARED ABSORPTION Seven methods are indicated for the preparation of previously dried test specimens and Reference Standards for analysis. The reference
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() Record the spectra of the test specimen and the corresponding USP Reference Standard over the range from about 2.6 µm to 15 µm (3800 cm–1 to 650 cm–1) unless otherwise specified in the individual monograph. The IR absorption spectrum of the preparation of the test specimen, previously dried under conditions specified for the corresponding Reference Standard unless otherwise specified, or unless the Reference Standard is to be used without drying, exhibits maxima only at the same wavelengths as that of a similar preparation of the corresponding USP Reference Standard.
Differences that may be observed in the spectra so obtained sometimes are attributed to the presence of polymorphs, which are not always acceptable (see Procedure under Spectrophotometry and Light-Scattering
![]() ![]() ULTRAVIOLET ABSORPTION The reference
![]() ![]() Dissolve a portion of the substance under examination in the designated Medium to obtain a test solution having the concentration specified in the monograph for Solution. Similarly prepare a Standard solution containing the corresponding USP Reference Standard.
Record and compare the spectra concomitantly obtained for the test solution and the Standard solution. Calculate absorptivities and/or absorbance ratios where these criteria are included in an individual monograph. Unless otherwise specified, absorbances indicated for these calculations are those measured at the maximum absorbance at about the wavelength specified in the individual monograph. Where the absorbance is to be measured at about the specified wavelength other than that of maximum absorbance, the abbreviations (min) and (sh) are used to indicate a minimum and shoulder, respectively, in an absorption spectrum. The requirements are met if the UV absorption spectra of the test solution and the Standard solution exhibit maxima and minima at the same wavelengths and absorptivities and/or absorbance ratios are within specified limits.
Auxiliary Information—
Please check for your question in the FAQs before contacting USP.
USP35–NF30 Page 133
Pharmacopeial Forum: Volume No. 36(2) Page 457
|