Change to read:
161
Change to read:
Endotoxin limit
(K × N)/V
K = limit for each device N = number of devices tested V = total volume of the extract
Kit:
Table 1. Selection of Product Units for Testing
Lambda ():
Product families:
Extracting or rinsing solution:
Sampling
Suitability testing (also known as test for interfering factors):
Apparatus:
Depyrogenate all glassware and other heat-stable materials in a hot air oven using a validated process. If employing plastic apparatus such as microplates and pipet tips for automatic pipetters, use apparatus that is shown to be free of detectable endotoxin and does not interfere with the test.
Analysts must be properly trained to perform the assay.
The assay sensitivity must be confirmed as described in 85 sections Test for Confirmation of Labeled Lysate Sensitivity for gel clot reagents and Assurance of Criteria for the Standard Curve for quantitative methods. Assay sensitivity confirmation must be carried out when a new batch of reagent (lysate or endotoxin) is used or when there is any change in the test conditions that could affect the outcome of the test.
Endotoxin limit:
(K × N)/V
K = amount of endotoxin allowed per device (20 Endotoxin Units per device; 2.15 for intrathecal devices) N = number of devices tested V = total volume of the extract or rinse
NoteThe volume of the rinse or extract may be adjusted for the size and configuration of the device.
The standard extracting fluid for extracting, rinsing, or soaking medical devices is Water for BET (see 85). If necessary, other solvents may be used after it has been demonstrated that they do not interfere with the performance of the assay. Either the entire device if claimed to be nonpyrogenic or all components of the device that are claimed to be nonpyrogenic, including fluid pathways, must come in contact with the extraction fluid for the entire course of the extraction.
The standard extraction method is to soak or immerse the device or flush the fluid pathway with extracting fluid that has been heated to 37 ± 1.0, keeping the extracting fluid in contact with the relevant surface(s) for NLT 1 h. Alternate extraction or rinsing methods may be used, but must be demonstrated to be equivalent or better than the standard method. Extracts from individual medical devices are typically pooled for testing.
Suitability:
For the suitability test, medical device extracts are considered to be the sample solutions under test and are tested as outlined in 85 for the chosen technique in the section Test for Interfering Factors.
If necessary, adjust the pH of the solution to be examined (or dilution thereof) so that the pH of the mixture of the lysate and sample solution falls within the pH range specified by the lysate manufacturer, usually pH 6.08.0. The pH may be adjusted by use of an acid, base, or suitable buffer as recommended by the lysate manufacturer. Acids and bases may be prepared from concentrates or solids with Water for BET (see 85) in containers free of detectable endotoxin. Buffers must be shown to be free of detectable endotoxin and interfering factors.
If the undiluted rinsing or extracting solution is determined to be unsuitable for 85, the test for interfering factors may be repeated utilizing dilution, or a method that will neutralize or eliminate the interfering substance.
If glucan interference is suspected, the use of a glucan blocking agent or endotoxin-specific lysate is permissible.
Dilution in Water for BET (see 85) or other validated solvent by a factor not to exceed the maximum valid dilution (MVD) is permissible. The MVD for medical devices is the following:
MVD = Endotoxin limit (of the rinsing or extracting solution in Endotoxin Units/mL)/
= sensitivity of the test method in Endotoxin Units/mL Interference may be overcome by suitable treatment of the device extract. Methods for overcoming interference and the qualification of such treatments are described in 85. To establish that the chosen treatment effectively eliminates interference without loss of endotoxins, perform the assay described in 85 using the preparation to be examined to which standard endotoxin has been added and which has then been submitted to the chosen treatment.
Changes to extraction techniques require a new suitability study. Because of possible changes in interference patterns between standard BET methods, a change in BET methodology (e.g., from gel clot limits test to kinetic chromogenic) requires a re-execution of the suitability study.
Routine testing:
Routine testing is to be performed as outlined in 85, following instructions for incubation and controls listed under the chosen assay technique.
Interpretation of test results:
If a device fails to meet this requirement, an investigation may be initiated per documented procedures. A failed, valid BET assay may not be repeated using Pyrogen Test 151 in its place.
Auxiliary Information
Please check for your question in the FAQs before contacting USP.
USP38NF33 Page 212
USP38NF33 Supplement : No. 1 Page 7039
Pharmacopeial Forum: Volume No. 40(3)
|