Propoxyphene Hydrochloride
(proe pox' i feen hye'' droe klor' ide).
DEFINITION
Propoxyphene Hydrochloride contains NLT 98.0% and NMT 102.0% of C22H29NO2·HCl, calculated on the dried basis.
IDENTIFICATION
• A. Infrared Absorption 197S
Sample solution:
50 mg/mL in chloroform
• B. Procedure
Sample solution:
17 mg/mL of Propoxyphene Hydrochloride in Purified Water
Analysis:
Treat 3 mL of Sample solution with 1 mL of 6 N ammonium hydroxide to precipitate the propoxyphene base. Filter to remove the precipitate, acidify the filtrate with 2 mL of nitric acid, and add 1 mL of silver nitrate TS.
Acceptance criteria:
A white, curdy precipitate that is soluble in an excess of 6 N ammonium hydroxide confirms the presence of silver chloride.
ASSAY
• Procedure
Solution A (0.02 M KH2PO4):
2.72 g/L monobasic potassium phosphate
Mobile phase:
Acetonitrile, triethylamine, and Solution A (53:0.1:47). Adjust with 85.0% phosphoric acid to a pH of 8.0. [NoteThe Mobile phase should be examined prior to use for particulates and clarity. ]
Standard solution:
0.11 mg/mL of USP Propoxyphene Hydrochloride RS in Mobile phase. Sonicate to dissolve if necessary.
Sample solution:
0.11 mg/mL of Propoxyphene Hydrochloride in Mobile phase. Sonicate to dissolve if necessary.
Chromatographic system
Mode:
LC
Detector:
UV 214 nm
Column:
4.6 mm × 15 cm; 3.5-µm packing L1
Flow rate:
1 mL/min
Injection size:
10 µL
System suitability
Sample:
Standard solution
Suitability requirements
Tailing factor:
NMT 1.5 for the propoxyphene hydrochloride peak
Relative standard deviation:
NMT 2.0%
Analysis
Samples:
Standard solution and Sample solution
Calculate the percentage of C22H29NO2·HCl in the portion of Propoxyphene Hydrochloride taken:
Result = (rU/rS) × (CS/CU) × 100
Acceptance criteria:
98.0%102.0% on the dried basis
IMPURITIES
Organic Impurities
• Procedure
Mobile phase:
Proceed as directed in the Assay.
Standard solution:
3 µg/mL of USP Propoxyphene Related Compound A RS, 2 µg/mL of USP Propoxyphene Related Compound B RS, and 1 µg/mL of USP Propoxyphene Hydrochloride RS in Mobile phase. Sonicate to dissolve if necessary.
Sample solution:
1.1 mg/mL of Propoxyphene Hydrochloride in Mobile phase. Sonicate to dissolve if necessary.
Chromatographic system:
Proceed as directed in the Assay, with the exception that the chromatographic run time is six times the retention time of propoxyphene.
System suitability
Sample:
Standard solution
Suitability requirements
Resolution:
NLT 2.0 between propoxyphene related compound A and propoxyphene related compound B, and NLT 2.0 between propoxyphene related compound B and propoxyphene
Relative standard deviation:
NMT 15%
Analysis
Samples:
Standard solution and Sample solution
Calculate the percentage of propoxyphene related compound A in the portion of Propoxyphene Hydrochloride taken:
Result = (rU/rS) × (CS/CU) × 100
Acceptance criteria:
See Impurity Table 1.
Calculate the percentage of propoxyphene related compound B as the hydrochloride in the portion of Propoxyphene Hydrochloride taken:
Result = (rU/rS) × (CS/CU) × (Mr1/Mr2) × 100
Acceptance criteria:
See Impurity Table 1.
Calculate the percentage of any other specified and unspecified impurity in the portion of Propoxyphene Hydrochloride taken:
Result = (rU/rS) × (CS/CU) × (1/F) × 100
Acceptance criteria:
See Impurity Table 1. [NoteDisregard any impurity less than 0.05%. ]
Impurity Table 1
SPECIFIC TESTS
• Optical Rotation, Specific Rotation 781S:
+52 to +57
Sample solution:
10 mg/mL in water, freshly prepared
• Loss on Drying 731:
Dry a sample at 105 for 3 h: it loses NMT 1.0% of its weight.
ADDITIONAL REQUIREMENTS
• Packaging and Storage:
Preserve in tight containers.
• USP Reference Standards 11
Auxiliary Information
Please check for your question in the FAQs before contacting USP.
USP38NF33 Page 5049
Pharmacopeial Forum: Volume No. 40(6)
|