Change to read:
![]() ![]() ![]() ![]()
Change to read:
![]() ![]() ![]() ![]()
Change to read:
• Method I.
![]() ![]()
Apparatus:
The determination may be carried out with
![]() ![]() ![]() ![]() ![]() ![]()
Procedure:
Fill the viscometer through tube (L) with a sufficient quantity of the sample liquid that is appropriate for the viscometer being used or by following the manufacturer’s instructions. Carry out the experiment with the tube in a vertical position. Fill bulb (A) with the liquid, and also ensure that the level of liquid in bulb (B) is below the exit to the ventilation tube (M). Immerse the viscometer in a water or oil bath stabilized at the temperature specified in the individual monograph, and control the temperature to ±0.1
![]() ![]() ![]() ![]() ![]() ![]()
Calibration:
Calibrate each viscometer at the test temperature by using fluids of known viscosities of appropriate viscosity standards to determine the viscometer constant, k. The viscosity values of the calibration standards should bracket the expected viscosity value of the sample liquid.
![]() ![]() Calculate the viscometer constant, k, in mm2/s2:
k =
![]() ![]()
Calculation of kinematic and Newtonian viscosities of sample fluid:
A capillary viscometer is chosen so that the flow time, t,
![]() ![]()
v = k × t
If the density of the fluid is known at the temperature of the viscosity measurement, then the Newtonian viscosity,
![]() ![]() ![]()
The flow time of the fluid under examination is the mean of NLT three consecutive determinations. The result is valid if the percentage of the relative standard deviation (%RSD) for the three readings is NMT 2.0%.
Change to read:
• Method II.
![]() ![]()
Apparatus:
The determination may be carried out with
![]() ![]() ![]()
Other viscometers, such as modified Ostwald-type capillary viscometers,1 may be used provided that the accuracy and precision is NLT that obtained with the viscometers described in this chapter.
![]()
Procedure:
Fill the tube with an amount of the sample that is appropriate for the viscometer being used or by following the manufacturer’s instructions. The volume of fluid used should be such that the lower bulb is not entirely emptied when the fluid is drawn up through the capillary tube to the uppermost graduation mark. Carry out the experiment with the tube in a vertical position. Immerse the viscometer in a water or oil bath stabilized at the temperature specified in the individual monograph, and control the temperature to ±0.1
![]()
Calibration and Calculation of kinematic and Newtonian viscosities of sample fluid:
Proceed as directed in Method I.
![]() ![]() ![]()
Auxiliary Information—
Please check for your question in the FAQs before contacting USP.
USP38–NF33 Page 679
Pharmacopeial Forum: Volume No. 39(5)
|